Reconfigurable I/O
NI 7831R User Manual
Reconfigurable I/O Devices for
PCI/PXI/CompactPCI Bus Computers
NI 7831R User Manual
April 2004 Edition
Part Number 370489B-01
Vista de página 0
1 2 3 4 5 6 ... 70 71

Resumo do Conteúdo

Página 1 - Reconfigurable I/O

Reconfigurable I/ONI 7831R User ManualReconfigurable I/O Devices for PCI/PXI/CompactPCI Bus ComputersNI 7831R User ManualApril 2004 EditionPart Number

Página 2

© National Instruments Corporation 1-1 NI 7831R User Manual1IntroductionThis chapter describes the NI 7831R, describes the concept of the Reconfigurab

Página 3 - Important Information

Chapter 1 IntroductionNI 7831R User Manual 1-2 ni.combetween devices. The NI PXI-7831R accesses the RTSI bus through the PXI trigger lines implemented

Página 4 - Compliance

Chapter 1 Introduction© National Instruments Corporation 1-3 NI 7831R User ManualOverview of Reconfigurable I/OThis section explains reconfigurable I/

Página 5 - Contents

Chapter 1 IntroductionNI 7831R User Manual 1-4 ni.comUser-Defined I/O ResourcesYou can create your own custom measurements using the fixed I/O resourc

Página 6

Chapter 1 Introduction© National Instruments Corporation 1-5 NI 7831R User ManualThe FPGA logic provides timing, triggering, processing, and custom I/

Página 7 - About This Manual

Chapter 1 IntroductionNI 7831R User Manual 1-6 ni.comYou can use Interactive Front Panel Communication to communicate directly with the VI running on

Página 8

Chapter 1 Introduction© National Instruments Corporation 1-7 NI 7831R User ManualCables and Optional EquipmentNational Instruments offers a variety of

Página 9 - Related Documentation

Chapter 1 IntroductionNI 7831R User Manual 1-8 ni.comRefer to Appendix B, Connecting I/O Signals, for more information about using these cables and ac

Página 10 - Introduction

Chapter 1 Introduction© National Instruments Corporation 1-9 NI 7831R User ManualSafety InformationThe following section contains important safety inf

Página 11 - Using PXI with CompactPCI

Chapter 1 IntroductionNI 7831R User Manual 1-10 ni.comOperate the NI 7831R at or below the installation category1 listed in the section Maximum workin

Página 12 - Flexible Functionality

SupportWorldwide Technical Support and Product Informationni.comNational Instruments Corporate Headquarters11500 North Mopac Expressway Austin, Texas

Página 13 - User-Defined I/O Resources

© National Instruments Corporation 2-1 NI 7831R User Manual2Hardware Overview of the NI 7831RThis chapter presents an overview of the hardware functio

Página 14 - Software Development

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-2 ni.comFigure 2-2. Parts Locator Diagram for the NI PXI-7831RSW1

Página 15 - LabVIEW Real-Time Module

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-3 NI 7831R User ManualFigure 2-3. Parts Locator Diagram for the NI PC

Página 16 - Cables and Optional Equipment

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-4 ni.comInput ModesThe NI 7831R input mode is software configurable. The input chann

Página 17 - Custom Cabling

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-5 NI 7831R User ManualConnecting Analog Input SignalsThe AI signals fo

Página 18 - Safety Information

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-6 ni.comFigure 2-4. NI 7831R Instrumentation AmplifierThe instrumentation amplifier

Página 19

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-7 NI 7831R User ManualTypes of Signal SourcesWhen configuring the inpu

Página 20 - Hardware Overview

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-8 ni.comFigure 2-5. Summary of Analog Input Connections+–+–V1AI<i>AISENSEAIGN

Página 21

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-9 NI 7831R User ManualDifferential Connection Considerations (DIFF Inp

Página 22 - Analog Input

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-10 ni.com Figure 2-6. Differential Input Connections for Ground-Referenced SignalsW

Página 23 - Input Range

Important InformationWarrantyThe NI 7831R is warranted against defects in materials and workmanship for a period of one year from the date of shipmen

Página 24

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-11 NI 7831R User Manual Figure 2-7. Differential Input Connections fo

Página 25

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-12 ni.comYou can fully balance the signal path by connecting another resistor of the

Página 26 - Types of Signal Sources

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-13 NI 7831R User ManualIn single-ended input modes, electrostatic and

Página 27

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-14 ni.compositive and negative inputs of the instrumentation amplifier. The instrume

Página 28 - Signal Sources

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-15 NI 7831R User ManualAnalog OutputThe NI 7831R has eight 16-bit AO c

Página 29 - Floating Signal Sources

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-16 ni.comFigure 2-10. Analog Output ConnectionsDigital I/OThe NI 7831R has 96 bidir

Página 30 - (see text)

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-17 NI 7831R User ManualTo interface to 5 V CMOS devices, you must prov

Página 31

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-18 ni.comFigure 2-11 shows signal connections for three typical DIO applications. Fi

Página 32 - Sources (NRSE Input Mode)

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-19 NI 7831R User Manualpaired with power or ground. Because the DIO li

Página 33

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-20 ni.comthe remaining devices on that PXI bus segment must have PXI/TRIG0 configure

Página 34 - Analog Output

ComplianceCompliance with FCC/Canada Radio Frequency Interference RegulationsDetermining FCC ClassThe Federal Communications Commission (FCC) has rul

Página 35 - Digital I/O

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-21 NI 7831R User ManualThe left local bus lines from the left peripher

Página 36

Chapter 2 Hardware Overview of the NI 7831RNI 7831R User Manual 2-22 ni.com4. Reinsert the NI 7831R into the PXI/CompactPCI chassis or PCI computer. R

Página 37

Chapter 2 Hardware Overview of the NI 7831R© National Instruments Corporation 2-23 NI 7831R User ManualField Wiring ConsiderationsEnvironmental noise

Página 38 - RTSI Trigger Bus

© National Instruments Corporation 3-1 NI 7831R User Manual3CalibrationCalibration is the process of determining and/or adjusting the accuracy of an i

Página 39

Chapter 3 CalibrationNI 7831R User Manual 3-2 ni.comIf you have NI-RIO installed, you can find the internal calibration utility at Start»All Programs»

Página 40 - Switch Settings

© National Instruments Corporation A-1 NI 7831R User ManualASpecificationsThis appendix lists the specifications of the NI 7831R. These specifications

Página 41 - Power Connections

Appendix A SpecificationsNI 7831R User Manual A-2 ni.comOvervoltage protection ...±42 VData transfers...

Página 42 - Field Wiring Considerations

Appendix A Specifications© National Instruments Corporation A-3 NI 7831R User ManualSettling TimeCrosstalk...

Página 43 - Calibration

Appendix A SpecificationsNI 7831R User Manual A-4 ni.comDC Transfer CharacteristicsINL...±0.5 L

Página 44 - External Calibration

Appendix A Specifications© National Instruments Corporation A-5 NI 7831R User ManualDigital logic levels Maximum output currentImax (source)...

Página 45 - Specifications

© National Instruments Corporation v NI 7831R User ManualContentsAbout This ManualConventions ...

Página 46 - Dynamic Characteristics

Appendix A SpecificationsNI 7831R User Manual A-6 ni.comPhase locked to PXI 10 MHzClock(NI PXI-7831R only) ...Adds 350 ps jitter, 300

Página 47 - Output Characteristics

Appendix A Specifications© National Instruments Corporation A-7 NI 7831R User Manualby the digital outputs use the following equation:Where j is the n

Página 48 - Voltage Output

Appendix A SpecificationsNI 7831R User Manual A-8 ni.comchassis, tested in accordance with IEC-60068-2-1 and IEC-60068-2-2Relative humidity range...

Página 49 - Reconfigurable FPGA

Appendix A Specifications© National Instruments Corporation A-9 NI 7831R User ManualSafetyThis product is designed to meet the requirements of the fol

Página 50 - Power Requirement

© National Instruments Corporation B-1 NI 7831R User ManualBConnecting I/O SignalsThis appendix describes how to make input and output signal connecti

Página 51 - Operating Environment

Appendix B Connecting I/O SignalsNI 7831R User Manual B-2 ni.comFigure B-2 shows the I/O connector pin assignments for the I/O connectors on the NI 78

Página 52 - Storage Environment

Appendix B Connecting I/O Signals© National Instruments Corporation B-3 NI 7831R User Manual. Caution Connections that exceed any of the maximum rati

Página 53 - CE Compliance

Appendix B Connecting I/O SignalsNI 7831R User Manual B-4 ni.com Connecting to CompactRIO Extension I/O ChassisYou can use the CompactRIO R Series Exp

Página 54 - Connecting I/O Signals

Appendix B Connecting I/O Signals© National Instruments Corporation B-5 NI 7831R User ManualConnecting to 5B and SSR Signal ConditioningNI provides ca

Página 55

Appendix B Connecting I/O SignalsNI 7831R User Manual B-6 ni.comFigure B-3. Connector Pinouts when Using NSC68-262650 CableThe NSC68-5050 cable conne

Página 56

ContentsNI 7831R User Manual vi ni.comSingle-Ended Connection Considerations ... 2-12Single-Ended C

Página 57

Appendix B Connecting I/O Signals© National Instruments Corporation B-7 NI 7831R User ManualNSC68-5050 cable header. In this case, you have access onl

Página 58

© National Instruments Corporation C-1 NI 7831R User ManualCUsing the SCB-68 Shielded Connector BlockThis appendix describes how to connect input and

Página 59

© National Instruments Corporation D-1 NI 7831R User ManualDTechnical Support and Professional ServicesVisit the following sections of the National In

Página 60 - DIO 24–39 Connector

Appendix D Technical Support and Professional ServicesNI 7831R User Manual D-2 ni.comIf you searched ni.com and could not find the answers you need, c

Página 61 - Shielded Connector Block

© National Instruments Corporation G-1 NI 7831R User ManualGlossarySymbol Prefix Valueppico10–12nnano10–9µ micro 10–6m milli 10–3k kilo 103Mmega106Ggi

Página 62 - Professional Services

GlossaryNI 7831R User Manual G-2 © National Instruments CorporationSquare root of.+5V +5 VDC source signal.AA Amperes.A/D Analog-to-digital.AC Alterna

Página 63

Glossary© National Instruments Corporation G-3 NI 7831R User ManualCCCelsius.CalDAC Calibration DAC.CH Channel—Pin or wire lead to which you apply or

Página 64 - Glossary

GlossaryNI 7831R User Manual G-4 © National Instruments CorporationDIO Digital input/output.DIO<i> Digital input/output channel signal.DMA Direc

Página 65

Glossary© National Instruments Corporation G-5 NI 7831R User ManualII/O Input/output—The transfer of data to/from a computer system involving communic

Página 66

GlossaryNI 7831R User Manual G-6 © National Instruments CorporationNnoise An undesirable electrical signal—Noise comes from external sources such as t

Página 67

© National Instruments Corporation vii NI 7831R User ManualAbout This ManualThis manual describes the electrical and mechanical aspects of the Nationa

Página 68

Glossary© National Instruments Corporation G-7 NI 7831R User ManualRRAM Random-access memory—The generic term for the read/write memory that is used i

Página 69

GlossaryNI 7831R User Manual G-8 © National Instruments CorporationTTHD Total harmonic distortion—The ratio of the total rms signal due to harmonic di

Página 70

About This ManualNI 7831R User Manual viii ni.comprograms, subprograms, subroutines, device names, functions, operations, variables, filenames, and ex

Página 71

About This Manual© National Instruments Corporation ix NI 7831R User ManualRelated DocumentationThe following documents contain information you might

Comentários a estes Manuais