National Instruments NI PXI/PCI-5114 Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Medição, teste e controlo National Instruments NI PXI/PCI-5114. National Instruments NI PXI/PCI-5114 User's Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 20
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 0
NI PXI/PCI-5114 Specifications
8-Bit 250 MS/s Digitizer
Unless otherwise noted, the following conditions were used for each
specification:
All filter settings
All impedance selections
Sample clock set to 250 MS/s
Typical values are representative of an average unit operating at room
temperature. Specifications are subject to change without notice. For the
most recent NI 5114 specifications, visit
ni.com/manuals.
To access the NI 5114 documentation, including the NI High-Speed
Digitizers Getting Started Guide, which contains functional descriptions of
the NI 5114 signals, navigate to Start»Programs»National Instruments»
NI-SCOPE»Documentation.
Hot Surface If the NI 5114 has been in use, it may exceed safe handling temperatures and
cause burns. Allow the NI 5114 to cool before removing it from the PXI chassis.
Contents
Vertical.................................................................................................... 2
Analog Input (Channel 0 and Channel 1) ........................................ 2
Horizontal................................................................................................ 6
Sample Clock................................................................................... 6
Phase-Locked Loop (PLL) Reference Clock................................... 7
CLK IN (Sample Clock and Reference Clock Input,
Front Panel Connector) ................................................................. 8
Trigger..................................................................................................... 8
Reference (Stop) Trigger ................................................................. 8
TRIG (External Trigger, Front Panel Connector)............................ 10
PFI 0 and PFI 1 (Programmable Function Interface,
AUX Front Panel Connectors) ............................................................. 10
TClk Specifications................................................................................. 12
Waveform Specifications........................................................................ 13
Calibration............................................................................................... 14
Vista de página 0
1 2 3 4 5 6 ... 19 20

Resumo do Conteúdo

Página 1 - Contents

NI PXI/PCI-5114 Specifications8-Bit 250 MS/s DigitizerUnless otherwise noted, the following conditions were used for each specification:• All filter s

Página 2 - Vertical

NI PXI/PCI-5114 Specifications 10 ni.comTRIG (External Trigger, Front Panel Connector)PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Pane

Página 3

© National Instruments Corporation 11 NI PXI/PCI-5114 SpecificationsAs an Input (Trigger)Destinations Start Trigger (Acquisition Arm)Reference (Stop)

Página 4

NI PXI/PCI-5114 Specifications 12 ni.comTClk SpecificationsNational Instruments TClk synchronization method and the NI-TClk driver are used to align t

Página 5

© National Instruments Corporation 13 NI PXI/PCI-5114 SpecificationsWaveform SpecificationsSpecification Value CommentsOnboard Memory Size8 MB per Cha

Página 6 - Horizontal

NI PXI/PCI-5114 Specifications 14 ni.comCalibrationPowerSpecification Val ue CommentsSelf-Calibration Self-calibration is done on software command. Th

Página 7

© National Instruments Corporation 15 NI PXI/PCI-5114 SpecificationsSoftwareSpecification Value CommentsDriver Software NI PXI-5114: NI-SCOPE 2.9 or l

Página 8 - Reference (Stop) Trigger

NI PXI/PCI-5114 Specifications 16 ni.comEnvironmentNI PXI-5114Note To ensure that the NI PXI-5114 cools effectively, follow the guidelines in the Mai

Página 9

© National Instruments Corporation 17 NI PXI/PCI-5114 SpecificationsNI PCI-5114Note To ensure that the NI PCI-5114 cools effectively, follow the guid

Página 10 - AUX Front Panel Connectors)

NI PXI/PCI-5114 Specifications 18 ni.comSafety, Electromagnetic Compatibility, and CE ComplianceSpecification Value CommentsSafety The NI 5114 meets t

Página 11

© National Instruments Corporation 19 NI PXI/PCI-5114 SpecificationsPhysicalSpecification Val ue CommentsDimensions NI PXI-5114 NI PCI-5114 —3U, One s

Página 12 - TClk Specifications

NI PXI/PCI-5114 Specifications 2 ni.comPower ...14

Página 13 - Waveform Specifications

National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on ni.com/legal f

Página 14 - Calibration

© National Instruments Corporation 3 NI PXI/PCI-5114 SpecificationsVoltage LevelsFull Scale (FS) Input Range and Programmable Vertical Offset50 Ω 1 MΩ

Página 15 - Software

NI PXI/PCI-5114 Specifications 4 ni.comFigure 1. NI 5114 Frequency Response (Typical)Bandwidth and Transient ResponseBandwidth (–3 dB)Range(Vpk-pk) B

Página 16 - Environment

© National Instruments Corporation 5 NI PXI/PCI-5114 SpecificationsSpecification Value CommentsSpectral CharacteristicsSpurious Free Dynamic Range wit

Página 17 - NI PCI-5114

NI PXI/PCI-5114 Specifications 6 ni.comHorizontalSample ClockSpecification Val ue CommentsSources Internal, Onboard Clock (internal VCXO)*External, CL

Página 18

© National Instruments Corporation 7 NI PXI/PCI-5114 SpecificationsPhase-Locked Loop (PLL) Reference ClockExternal Sample ClockSources CLK IN (front p

Página 19 - Physical

NI PXI/PCI-5114 Specifications 8 ni.comCLK IN (Sample Clock and Reference Clock Input, Front Panel Connector)TriggerReference (Stop) TriggerSpecificat

Página 20 - Where to Go for Support

© National Instruments Corporation 9 NI PXI/PCI-5114 SpecificationsAnalog Trigger (Edge, Window, and Hysteresis Trigger Types)Sources CH 0 (front pane

Comentários a estes Manuais

Sem comentários