SCXI-1125 Calibration Procedure 10 ni.com
Verifying Adjusted Values
After completing the adjustments to your SCXI-1125 module, it is
important that you verify the analog input operation by repeating the steps
listed in the Verifying Gain and Offset Errors section of this document.
Reverifying after adjustments ensures that your SCXI-1125 module is now
operating within its specifications.